U type coriolis mass flowmeter installation suggestions
1.Stay as far away from the vibration of source as possible
2.Fix well but the U-shape should be fixed at both ends of the flange. Do not touch the U-shape
3.No straight pipe length limit
4.Direction
Suggest to install upwards downwards or side installation ,parallel to the ground or tilted are all not suggested Detais please find as bellow:
Liquid:
The meter should be installed downward or sideward when measuring liquid flow so that air cannot get trapped inside the tubes.
Gas:
The meter should be installed upward when measuring gas flow , so that liquid cannot get trapped inside the tubes
Turbid liquid
The meter should be installed sideward or upward to avoid particulate matter accumulated in the measuring tube. The flow direction of medium goes from the bottom up through the sensor.