Measuring Principle | Hydrostatic | |
Characteristic |
Cable probe for hydrostatic level measurement, with ceramic cell Diameter = 22mm: General applications, Drinking Water Diameter = 42mm: Wastewater applications Diameter = 29mm: Salt water applications Communication via HART |
|
Supply voltage |
10.5 ...35V DC |
|
Reference Accuracy |
0.2% Option: 0.1% |
|
Long term stability | 0.1% of URL/year | |
Process temperature |
-10°C...70°C (14°F...158°F) Diameter 29mm (1.14"): 0°C...50°C (32°F...122°F) |
|
Ambient temperature |
-10°C...70°C (14°F...158°F) Diameter 29mm (1.14"): 0°C...50°C (32°F...122°F) |
|
Measuring cell |
100mbar...20bar (1.5psi...300psi) |
|
Max. overpressure limit |
40 bar (580psi) Diameter = 29mm: 25 bar (362psi) |
|
Process connection |
Suspension clamp Cable mounting screws |
|
Output |
4...20 mA HART |
|
Certificates / Approvals |
ATEX, FM, CSA, CSA C/US, IEC Ex, TIIS, INMETRO, NEPSI GL/ ABS/ LR/ BV/ DNV Marine NSF/ KTW/ ACS drinking water |
|
Options |
Absolute measuring cell Integrated temperature sensor Pt100, 4-wire Customer specific adjustment Customer specific cable marking |