80I Coriolis flowmeter
80I Coriolis flowmeter
Name: 80I Coriolis flowmeter
The straight single-tube flowmeter with a compact or remote transmitter
Description
Measuring Principle Coriolis
Product headline The straight single‐tube flowmeter with a compact or remote transmitter. Measuring liquids and gases in applications requiring low pressure loss and gentle fluid treatment.
Sensor features Energy‐saving – full bore design enables minimal pressure loss. Fewer process measuring points – multivariable measurement (flow, density, temperature). Space‐saving installation – no in/outlet run needs. Straight, easy cleanable single‐tube system. TMB® technology.
Transmitter features Cost‐effective – dedicated design for standard applications. Safe operation – display provides easy readable process information. Fully industry compliant – IEC/EN/NAMUR. 2‐line backlit display with push buttons. Device in compact or remote version.
Nominal diameter range DN 8 to 80 (⅜ to 3")
Wetted materials Measuring tube: Titanium grade 9
Connection: Titanium grade 2
Measured variables Mass flow, density, temperature, volume flow, corrected volume flow, reference density
Max. measured error Mass flow (liquid): ±0.15 %
Volume flow (liquid): ±0.15 %
Mass flow (gas): ±0.5 %
Density (liquid): ±0.0005 g/cm³
Measuring range 0 to 180 000 kg/h (0 to 6600 lb/min)
Max. process pressure PN 100, Class 600, 63K
Medium temperature range –50 to +150 °C (–58 to +302 °F)
Ambient temperature range Standard: –20 to +60 °C (–4 to +140 °F)
Option: –40 to +60 °C (–40 to +140 °F)
Sensor housing material 1.4301/1.4307 (304L), corrosion resistant
Transmitter housing material Powder‐coated die‐cast aluminium
1.4301 (304), sheet
CF3M (316L), cast
Degree of protection IP67, type 4X enclosure. Remote transmitter: IP67, type 4X enclosure
Display/Operation 2‐line backlit display with push buttons
Configuration via local display and operating tools possible
Outputs 3 outputs:
0‐20 mA (active)/4‐20 mA (active/passive)
Pulse/frequency/switch output (passive)
Inputs Status input
Digital communication HART
PROFIBUS PA
Power supply DC 16 to 62 V
AC 85 to 260 V (45 to 65 Hz)
AC 20 to 55 V (45 to 65 Hz)
Hazardous area approvals ATEX, IECEx, FM, CSA, NEPSI, TIIS
Other approvals and certificates 3.1 material, calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025), NAMUR, SIL
PED, CRN
3A, EHEDG