Measuring Principle | Hydrostatic | |
Characteristic | Smart and reliable pressure transmitter, with piezoresistive measuring cell and metal welded process isolating diaphragm | |
Supply voltage |
4...20mA HART 11,5...45V DC (Non Ex): Ex ia: 11,5...30V DC PROFIBUS PA: 9...32 V DC (Non Ex) FOUNDATION Fieldbus: 9...32 V DC (Non Ex) |
|
Reference Accuracy |
Standard 0,15% Platinum 0,075% |
|
Long term stability |
< 0,1% of URL/year < 0,175% of URL/ 5 years |
|
Process temperature |
-40°C...130°C (-40°F...266°F 150°C for 1h (302°F for 1h) |
|
Ambient temperature |
-40°C...85°C (-40°F...185°F) |
|
Measuring cell |
400mbar...400bar (15psi...6000psi) relative/ absolute |
|
Smallest calibratable span | 20 mbar (2.9 psi) rel./ abs. | |
Vacuum resistance |
10 mbar (1.45 psi) abs. |
|
Max. Turn down | 20:1 | |
Max. overpressure limit | 600bar (9000psi) | |
Process connection |
Thread: G1/2...G2, MNPT1/2...MNPT2 Flange: DN25...DN80, ASME 1 1/2"...4", |
|
Hygienic |
Tri-Clamp DIN11851 NEUMO Varivent N SMS DRD |
|
Material process membrane |
316L, AlloyC, Tantal, Rhodium>Gold |
|
Material gasket | None, measuring cell welded | |
Fill fluid |
Silicone oil, Inert oil |
|
Material housing |
Die-cast aluminum, AISI 316L |
|
Output |
4...20 mA 4...20 mA HART PROFIBUS PA FOUNDATION Fieldbus |
|
Certificates / Approvals |
ATEX, FM, CSA, CSA C/US, IEC Ex, TIIS, INMETRO, NEPSI SIL EN10204-3.1 NACE 3A, EHEDG CoC ASME-BPE GL/ ABS/ LR/ BV/ DNV Marine NSF Drinking water |
|
Options |
Display PWIS free |