Measuring Principle | Hydrostatic | |
Characteristic |
Digital transmitter with piezoresistive sensor and diaphragm seal Modular transmitter Long term stability Minimum oil volume process connection Enhanced safety via self diagnostic functions Secondary process barrier |
|
Supply voltage |
4...20mA HART 10,5...45V DC (Non Ex): Ex ia: 10,5...30V DC PROFIBUS PA: 9...32 V DC (Non Ex) FOUNDATION Fieldbus: 9...32 V DC (Non Ex) |
|
Reference Accuracy | 0,075% + influence of diaphragm seal | |
Long term stability | 0,05% of URL/year | |
Process temperature |
-70°C...400°C (-94°F...752°F) |
|
Ambient temperature |
-40°C...85°C (-40°F...185°F) |
|
Measuring cell |
100mbar...700bar (1.5psi...10.150psi) relative/ absolute |
|
Smallest calibratable span | 5 mbar (0.075 psi) | |
Vacuum resistance | 10 mbar (1.45 psi) | |
Max. Turn down | 100:1 | |
Max. overpressure limit | 1050bar (15.750psi) | |
Process connection |
Thread: G1/2...G2, R1/2, MNPT1/2...MNPT2, NPT1/2...NPT1 Flange: DN25...DN100, ASME 1"...4", JIS 10K Diaphragm seal |
|
Hygienic |
Tri-Clamp DIN11851 NEUMO Varivent SMS DRD Universal adapter |
|
Material process membrane |
316L, AlloyC, Tantal Rhodium> Gold PTFE |
|
Material gasket | None, diaphragm welded | |
Fill fluid |
Silicone oil, Inert oil, Vegetable oil, High temperature oil, Low temperature oil, |
|
Material housing |
Die-cast aluminum, AISI 316L |
|
Output |
4...20mA HART PROFIBUS PA FOUNDATION Fieldbus |
|
Certificates / Approvals |
ATEX, FM, CSA, CSA C/US, IEC Ex, TIIS, INMETRO, NEPSI Overfill prevention WHG SIL EN10204-3.1 NACE 3A, EHEDG GL/ ABS Marine |
|
Options |
HistoROM/M-Dat 4-line digital display SS- or Aluminiumhousing Separate housing |
|
Specialities |
Diagnostic functions TempC Membrane |